VITA based Neutron Source - Status and Prospects V. Aleynik, A. Bashkirtsev, M. Kamkin, D. Kasatov, A. Kuznetsov, A. Makarov, I. Shchudlo, I. Sorokin, S. Taskaev , M. Tiunov Budker Institute of Nuclear Physics, Novosibirsk, Russia XXIII Russian Particle Accelerator Conference September 24-28, 2012, Saint Petersburg, Russia

Transcript of VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per...

Page 1: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

VITA based Neutron Source - Status and Prospects

V. Aleynik, A. Bashkirtsev, M. Kamkin, D. Kasatov, A. Kuznetsov, A. Makarov, I. Shchudlo, I. Sorokin,

S. Taskaev, M. Tiunov

Budker Institute of Nuclear Physics, Novosibirsk, Russia

XXIII Russian Particle Accelerator ConferenceSeptember 24-28, 2012, Saint Petersburg, Russia

Page 2: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

VITA team

Page 3: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

VITA team

Page 4: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

MondayIgor Sorokin – X-ray Radiation of the High-Voltage Elements of the Tandem-

Accelerator With Vacuum Insulation

WednesdayAlexander Kuznetsov – Calibration Testing of the Stripping Target of the Vacuum Insulated

Tandem AcceleratorVladimir Aleynik – The Control System for the Tandem-BNCT Accelerator Facility

Alexander Makarov – Optimization of the Negative Hydrogen Ion Beam Injection into the Tandem Accelerator with Vacuum Insulation

Page 5: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

boron neutron capture therapy of cancer

epithermal neutrons: 0,5 eV – 10 keV109 cm–2 c–1

accelerators: > 2MeV, > 10 mA

Page 6: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

dc source of negative hydrogen ions(8 mA 21 keV)

Vacuum InsulationTandem Accelerator(dc 5 mA 2.5 MeV protons)

Lithium target for 7Li(p,n)7Be

High voltage power supply

dose rate in tumor ~ 1 RBE Gy per minute

Page 7: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

Page 8: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Introduction

Facility for development of BNCT in operation since 2008

Page 9: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Novel accelerator (VITA) has been proposed and createdBayanov et al. NIM A 413 (1998) 397Sorokin and Shirokov. Instr. Exper. Techniques 50 (2007) 719Kudryavtsev et al. Rev. Sci. Instr. 79 (2008) 02C709

Lithium target (solid, thin, metallic) has been developed, manufactured and optimizedBayanov et al. Applied Radiation and Isotopes 61 (2004) 817Bayanov et al. Journal of Physics 41 (2006) 460Bayanov et al. Instr. Exper. Techniques 51 (2008) 147 Bayanov et al. Instr. Exper. Techniques 51 (2008) 438Astrelin et al. J. Nuclear Materials 396 (2010) 43Bayanov et al. Instr. Exper. Techniques 53 (2010) 883Kandiev et al. Applied Radiation and Isotopes 69 (2011) 1632

Neutron generation was realizedKuznetsov et al. Techn. Phys. Lett. 35 (2009) 1

First in vitro investigations were carried outMostovich et al. Bull. Experimental Biology and Medicine 151 (2011) 264

Introduction

Page 10: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Our main problems were:

1. Low reliability of feedthrough insulatorIt required re-assembly every year because of SF6 leakage through vacuum seal (Indium wire).Re-assembly takes 3-4 months.

We have changed Indium wire by vacuum rubber this summer.Result is still positive.

Page 11: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Our main problems were:

2. Low proton current We obtained 3 mA proton beam during several seconds,but average current was only 0.1 – 0.7 mA @ 2008-2010.

We have carried out several experiments to increase current

Page 12: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Dark current:

• The accelerator has a high electric field in the electrode gap – about 25 kV/cm, and a large total area of the electrodes – tens of square meters.

In such a system dark current must inevitably occur, which may have a significant impact on the potential distribution along the accelerating channel.

Dark current was investigated.

Facility was modified to prevent the occurrence of high intensity dark current

Приборы и техника эксперимента (направлено)

Page 13: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Electrostatic lens :

• The accelerator has strong entry electrostatic lens.

We installed a detector in entry lens area and measured beam profile depending on the focusing magnetic lenses current

We determined:1. full compensation of the space charge

takes place in the transport channel 2. transverse ion temperature is equal to 1 eV

at the plasma boundary of the ion source

This study resulted in better focusing of the beam required for acceleration of the beam without significant losses.

22-channel detectormagneticlenses

Page 14: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Stripper:

We installed new gas stripper (400 mm, ∅ 16 mm) and measured output current depending on stripping target density

We determined:1. The dependence of the output current on stripping

target density is in good agreement with the calculated one when injected H– current is around hundreds of microamps

2. when the injected current is around several milliamps there is some reduction of the output current in the saturation region with increasing gas supply

-150

-100

-50

0

50

100

150

200

0 0,0005 0,001 0,0015 0,002 0,0025 0,003

Page 15: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Main result of these accelerator investigations:

These investigations allowed us to move to a long stable operation with much higher average current – 1.5 – 2 mA instead of 0.1 – 0.7 mAachieved previously

Page 16: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Beam Shaping Assembly for therapeutic epithermal neutron beam:

2 – 3 MeV proton beams and Li targets:

Italy team Palamara et al. 10 ICNCT (2002)Ohio Un. team Khorsandi and Blue. 11 ICNCT (2004)Obninsk team Kononov et al. 11 ICNCT (2004)KURRI team Kobayashi et al. 12 ICNCT (2006)IBA team Jongen et al. 12 ICNCT (2006)Argentina team Kreiner et al. 14 ICNCT (2010)

Dose rate @ 10 mA ~ 1 RBE Gy per minuteAdvanced depth ~ 6 – 10 cmTherapeutic ratio ~ 4 – 6

Page 17: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Standard mode @ 2.5 MeV

Dose rate @ 10 mA =1.5 RBE Gy / min

AD = 11 cm

TR = 5

0

5

10

15

20

25

30

35

40

1,E+00 1,E+01 1,E+02 1,E+03 1,E+04 1,E+05

Page 18: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

0,00

0,05

0,10

0,15

0,20

0,25

0,30

0,35

0,40

0 5 10 15

52.5 ppm 10-B14N(n,p)14Celastic scatteringgamma

Current status

Standard mode @ 2.5 MeV

Dose rate @ 10 mA =1.5 RBE Gy / min

AD = 11 cm

TR = 5

0,00

0,20

0,40

0,60

0,80

1,00

1,20

1,40

1,60

0 5 10 15

tumornormal tissue

Gy

per m

inut

e @

10

mA

RB

E G

ype

r min

ute

@ 1

0 m

A

Page 19: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

advantages:1. low yield of

neutrons2. low activation of the

lithium target and the facility

Near threshold mode @ 1.95 MeV

Dose rate @ 10 mA =1 RBE Gy / min

AD = 7 cm

TR = 2.5

0

5

10

15

20

25

30

35

40

1,E+00 1,E+01 1,E+02 1,E+03 1,E+04 1,E+05

Page 20: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

Orthogonal mode @ 2.5 MeV

Dose rate @ 10 mA =3 RBE Gy / min

AD = 12 cm

TR = 4

advantage:1. high dose rate2. à la gantry possibility

0

5

10

15

20

25

30

35

40

1,E+00 1,E+01 1,E+02 1,E+03 1,E+04 1,E+05

Page 21: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

current design of the facility

Current status

Page 22: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Current status

detectorfor proton beam test

Orthogonal modeand rotating BSAallows to useany angle neutron beam

Page 23: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Prospects

BNCT:

• VITA improving:new H– source – 15 mA 150 kVnew oil-free vacuum pumping

• Li target and BSA development

• neutron spectra and flux measurements

• blistering and target investigations

• in vitro and in vivo investigation

• new design of medical facility

Page 24: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Prospects

other applications:

• generation of 9.17 MeV γ-rays @ 1.747 MeV protonsfor development of technique of monochromatic γ-rays resonance absorption in nitrogen to detect explosives and narcotics

• measuring the cross section and the spectrum of α-particles from neutronless thermonuclear reaction 11B(p,α)αα

• dating of rock formation (apatite) by inducing the fission of uranium contained in the rock

• forming monochromatic beam of epithermal neutrons for calibration of dark matter detector

• ____________

Page 25: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Conclusion

Conclusion:

VITA based facility for development of BNCT is now in use @ 1 - 2 mA

Page 26: VITA based Neutron Source - Status and Prospects...normal tissue Gy per minute @ 10 mA RBE Gy per minute @ 10 mA Current status advantages: 1. low yield of neutrons 2. low activation

Thank youfor your attention!

www.inp.nsk.su/bnct/