© CEA 2007. Tous droits réservés. Toute reproduction totale ou partielle sur quelque support que ce soit ou utilisation du contenu de ce document est interdite sans l’autorisation écrite préalable du CEAAll rights reserved. Any reproduction in whole or in part on any medium or use of the information contained herein is prohibited without the prior written consent of CEA
2007
www.leti.fr
FIB/SEM Dual beam :FEI Expida 1285300 mm FOUP & 200 mm Open CassetteSEM column : Sirion (3 nm resolution)FIB column : Sidewinder (low kV thinning)Beam Chemistries:
Tungsten deposition SiO2 deposition Insulator enhanced etch
Nanolift option (TEM sample preparation & transfer): ChunkWizard - Omniprobe - TSU
X-Ray Analysis (Oxford)
TEM :FEI Tecnai G² F20 STWIN TMPVoltage : 200 kVElectron Source : FEG
TEM point resolution : 0.24 nm TEM line resolution : 0.14 nm STEM resolution : 0.20 nm
2kx2k GATAN CCDX-Ray Analysis (EDAX)
9 - First TEM Observation00:30; 20%
1 - Wafer loading (inclunding : area of Interest
localization and electron assisted W deposition
protective layer)
00:15; 10%
2 - Sample milling(5x5 µm)
00:15; 10%
3 - Lift out(sample extraction)
00:15; 10%
4 - Capsule Loading 00:05; 3%
5 - Sample transfer to TEM grid00:10; 7%
6 - Thinning00:40; 27%
7 - Capsule Unloading00:05; 3%
8 -Transfer to TEM 00:15; 10%
In-line transmission electron microscopy for micro and nanotechnologies R&D
V. Delaye, F. Andrieu, F. Aussenac, O. Faynot, R. Truche, C. Carabasse, A. L. Foucher, A. DanelCEA-LETI, MINATEC, 17 rue des Martyrs, 38054 Grenoble Cedex 9 France
Phone: +33 (0)4 38 78 40 08 e-mail: vincent.delaye@cea.fr
IntroductionAdvanced microelectronics, micro and nanotechnologies characterizations needs :
Reduced cycle time (cost) High resolution imaging and analysis of materials and structures with
nanometers dimensionsAn adapted tool is Transmission electron microscope (TEM), but :
Dedicated for off-line laboratories Destructive Delicate and time consuming sample preparation
To meet the needs of such R&D facility, as much in term of cycle time as resolution :
Full 200 and 300 mm wafer FIB-SEM dual beam system for sample preparation (100 nm thick lamella)
200 kV TEM installed close to dual beam and inside cleanroom.
2h30 to 6h00 hours/sample
200 & 300 mm wafers
TEM samples
In-line TEM resolutionThe microscope supplier (FEI) specifications are all achieved in dedicated area except for 50Hz (60dB instead of less than 54 dB) acoustic level. However : Resolution tests in TEM and STEM modes have been successfully conduced with the FEI Tecnai S-Twin 200 kV microscope installed.
Preparation to observation flowStarting from wafer loading in the dual beam to TEM picture delivery : A cycle time between 2h30 and 6h00 hours has been obtained depending on the sample size and the lamella thickness required.
This flow includes : Electron beam assisted tungsten protective layer deposition to reduce
surface damage to 1 nm instead of 30 nm with ion beam Low-kV (5kV) ion beam final thinning for less than 100 nm lamellas
(nanometer gate oxide measurement for example)
Wafer return for front-end levelsAn other major concern in R&D and manufacturing semiconductor industry is to reduce wafer costs due to destructive characterization. It has been already shown that sample preparation impact on wafer is limited to 1 mm around the FIB crater; such impact allows the industry (1) to re-introduce wafers after TEM sample extraction at back-end levels.
Dedicated TEM area inside the cleanroom : Anti-vibration TMC table Acoustical walls Adapted lighting conditions Adapted air conditioning flows Magnetic field canceling system
Time management for
a 5x5x0.1µm lamella
NMOSL1B
PTGL250
To go further, we made 2 samples extractions (common gate pattern NMOSL1B L=70nm and isolated transistor PTGL25 W=50nm) at 3 front-end process levels on two FDSOI (Fully Depleted SOI) wafers to study the impact on electrical performances.
The device extraction does not affect the electrical performance of all the devices further than 500 µm (from another pattern in the same die). On NMOSL1B common gate pattern, devices of the whole pattern are completely scrapped. On PTGL250 isolated transistors other devices, from 180 to 400 µm, on the same pattern work (even if, with degraded performance) except for Step 2.
This new result confirms the localized impact of a FIB extraction around the crater for advanced front-end levels.
Wafer mapping
(11 e-beam dies)
Die description with
the 2 localized patterns
SEM image from isolated PTGL250
transistors pattern with FIB crater zoomElectrical tests results
-0,54
-0,52
-0,5
-0,48
-0,46
-0,44
-0,42
-0,4
-0,38
-0,36
-0,34
VT
(V
)
Max
Min
VT
D01
D03
D04
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
Gm
(µ
S/µ
m)
Max
Min
gm
D01
D03
D04
0.14 nm line resolution
HR-TEM Si 110 HR-STEM Si 110
Chunk milling Probe welding Chunk lift-out FIB crater
Transfer to grid Chunk approach
Sample preparation to observation flow example
Chunk view Thinned chunk
FIB sample preparation
Sample
transfer
TEM analysis
Inline TEM Dedicated Cleanroom Area
200 & 300 mm
Lots wafers
TEM sample preparation
using FIB/SEM
DualBeam
TEM/STEM sample imaging and EDX(1(1stst analysis) analysis)
D01
D02 D03
D04D05
D06
D07
D08
D09
D10
D11
D01
D02 D03
D04D05
D06
D07
D08
D09
D10
D11
W=0.07µm300 µm from W=0.04µm
W=0.10µm405 µm from W=0.04µm
W=0.15µm360 µm from W=0.04µm
W=0.25µm405 µm from W=0.04µm
W=0.03µm180 µm from W=0.04µm
W=0.04µmSite location
W=0.05µm180 µm from W=0.05µm
W=0.07µm300 µm from W=0.04µm
W=0.10µm405 µm from W=0.04µm
W=0.15µm360 µm from W=0.04µm
W=0.25µm405 µm from W=0.04µm
W=0.03µm180 µm from W=0.04µm
W=0.04µmSite location
W=0.05µm180 µm from W=0.05µm
PTGL250 transistors : Voltage threshold
PTGL250 transistors : Transconductance
FDSOI Transistor
FIB/SEM TEM
Metal gate
Ste
p #
Lev
el
Des
cri
pti
on
Nu
mb
er o
f L
ift-
ou
t
Waf
er
(slo
t )
Fro
m a
12
waf
ers
lot
Die
(s)
Tra
ns
isto
r
Ele
ctri
cal
test
(o
ther
dev
ice
s ar
ea a
nd
die
s)
Ele
ctri
cal
test
(s
ame
dev
ices
ar
ea)
PTGL250 Vt Step 1
After active area patterning
2 11 01 & 03 NMOSL1B
PTGL250 Step 2
During gate stack formation
2 12 01 NMOSL1B
PTGL250 Vt Step 3
Before extension formation
2 12 04 NMOSL1B
> 5
00µ
m
< 5
00µ
m
Conclusion : No measured impact die to die Reduced impact within a pattern, but depending on following process step
200 and 300 mm wafers can be re-introduced within the process flow
Applications : step by step analysis, process monitoring, defectivity.